Sabtu, 11 Mei 2013

PAL SYSTEM TELEVISION MEASUREMENT : CHROMINANCE NONLINEAR GAIN

Chrominance Nonlinear Gain

DEFINITION
Chrominance nonlinear gain distortion is present when chrominance gain is affected by chrominance amplitude. These amplitude errors are a result of the system's inability to uniformly process all amplitudes of
chrominance information.
Chrominance nonlinear gain distortion is the amplitude deviation expressed as a percentage of the nominal amplitude. This measurement is made on the lowest and highest chrominance levels with the middle level
normalized to its nominal value.
The larger of the two resulting numbers is generally taken as the overall result.
This distortion should be measured at different average picture levels and the worst distortion should be quoted.
PICTURE EFFECTS
Chrominance nonlinear gain is often seen as attenuation of relatively high amplitude chrominance signals. It will appear in the TV picture as incorrect colour saturation.
TEST SIGNAL
A modulated pedestal signal, sometimes called a three level chrominance bar, is used to measure this distortion. This signal consists of a single phase, three level chrominance packet superimposed on a constant luminance level. A typical modulated pedestal signal will have a 350 mV luminance level and 140, 420, and 700 mV chrominance levels. This signal element is sometimes part of combination signals used as ITS.
Figure 86. A modulated pedestal test signal.
MEASUREMENT METHODS
Chrominance nonlinear gain distortion is quantified by measuring how much the amplitudes of  the chrominance packets deviate from their nominal values.
Waveform Monitor.
The waveform monitor graticule should be used for this measurement. First use the waveform monitor variable gain to normalize the middle subcarrier packet to its prescribed value of 420 mV. The amount of chrominance nonlinear gain distortion is the largest deviation from nominal value for the other two packets expressed as a percentage of the nominal amplitude of the affected packet.

Figure 87. This signal exhibits chrominance nonlinear gain distortion. Note that the amplitude of the largest packet is reduced.

VM700T Automatic Measurement.
Select CHROMINANCE NONLINEARITY in the VM700T MEASURE menu to make this Measurement. Chrominance nonlinear gain is shown on the upper graph. This parameter can also be measured in the VM700T AUTO mode.
Figure 88. The VM700T Chrominance Nonlinearity display.
NOTES
23. Chroma Filter.
It is sometimes recommended that waveform monitor chroma filter be enabled when measuring chrominance
nonlinear gain. While the chroma filter will make the display more symmetrical, the same results should be obtained either way since it is the peak-to-peak amplitudes being measured.
A possible exception is a case where chrominance harmonic distortion is present. The chrominance filter can remove the effects of harmonic distortion which are likely to be different for each chrominance level.


Lihat juga

Table of Contents
Preface                               
3
4
EQUIPMENT REQUIREMENTS         
4
CALIBRATION                      
6
6
DEMODULATED RF SIGNALS         
8
TERMINATION                     
8
8
PERFORMANCE GOALS              
8
9
9
10
12
SCH Phase                    
15
II LINEAR DISTORTIONS         
18
Chrominance-to-Luminance Gain and Delay                
19
Short Time Distortion           
24
Line Time Distortion            
26
Field Time Distortion           
28
Long Time Distortion        
30
Frequency Response            
31
Group Delay                    
36
K Factor Ratings                 
38
41
Differential Phase           
42
Differential Gain              
46
50
5 2
5 3
54
55
56
57
Signal-to-Noise Ratio             
58
6 0
61
63
64
APPENDICES

67



Tidak ada komentar:

Posting Komentar

ucx','_assdop');